FULLTEST 3
EN - 12
4.3. MEASUREMENT FUNCTION SELECTION
Press the FUNC hard key (yellow), the function selection screen will appear.
R
PE-4WIRE
R
PE-2WIRE
LOOP
RCD
PHASE
SEQ
I
CLAMP
I
LEAK
POWER
M
DIELECT
R
A
U
RES
Figure 3: Function selection screen
Select wished function by pressing appropriate touch-screen key, basic measurement screen of
selected function will appear, see an example of RPE-2WIRE basic measurement screen below. Other
functions use adapted screens, but follow the same system.
Figure 4: Basic measurement screen in RPE-2WIRE function
1 ..... Selected function.
2 ..... Progress bar, it follows measurement time during the measurement (in TIMER mode only).
3 ..... Two lines reserved for sub-results (no value in basic screen yet).
4 ..... Measurement screen touch-screen key.
5 ..... Measurement parameters touch-screen keys.
6 ..... Test lead calibration status (LEADS CALIBATED or LEADS NOT CALIBRATED).
7 ..... Measurement result field (result in green colour - OK, result in red colour - not OK, result in white
colour – not judged).
8 ..... Real time clock (hh.mm.ss).
9 ..... Unit of the measurement result.
10 ... Set measurement time (in TIMER mode only).