The following keys are active on this page:
The F1 key (or the PAG item on the display) advances to the following page of
saved values.
The ESC key (or the smart icon on the display) to exit the function and go back
to the “Recording analysis” page (Fig. 143).
In this page, the following symbols are used:
Patt System total Active Power
Preatt Total Reactive Power
Papp Total Apparent Power
Pf Total Power Factor
CosPhi System total CosPhi
CosPhi represents the theoretical limit value which
can be reached by the Power factor if all harmonics
are eliminated from the electrical system. For
dimensioning a power factor correction system,
reference is usually made to the CosPhi parameter
value.
Fig. 180: Page 2/5 of numeric values for three-phase 4-wire system
The following keys are active on this page:
The F1 key (or the PAG item on the display) advances to the following page of
saved values.
The ESC key (or the smart icon on the display) to exit the function and go back
to the “Recording analysis” page (Fig. 143).
In this page, the following symbols are used:
V1N Phase L1 - Neutral Voltage
I1 Phase L1 Current
Patt1 Phase L1 Active Power
Preatt1 Phase L1 Reactive Power
Papp1 Phase L1 Apparent Power
Pf1 Phase L1 Power Factor
CosPhi1 Cosine of the Phase delay between the Phase L1
Voltage and Current fundamentals
CosPhi represents the theoretical limit value which
can be reached by the Power factor if all harmonics
are eliminated from the electrical system. For
dimensioning a power factor correction system,
reference is usually made to the CosPhi parameter
value.
Fig. 181: Page 3/5 of numeric values for three-phase 4-wire system
The following keys are active on this page:
The F1 key (or the PAG item on the display) advances to the following page of
saved values.
The ESC key (or the smart icon on the display) to exit the function and go back
to the “Recording analysis” page (Fig. 143).